Model PI-700

PI-700 Instruction Manual

Rev. 2.5

Page 11 of 54

 Detcon Span Gas (See Detcon for Ordering Information). Recommended span gas is 50% of range

with Isobutylene in air or N

2

balance.

 Detcon P/N 985-241100-321

In-Line Humidifying Tube 24"

a) Attach the calibration adapter to the threaded sensor housing or connect tubing to integral cal port. Apply

the test gas at a controlled flow rate of 200 - 500cc/min using the in-line humidifying tube, (200cc/min is
the recommended flow).

Observe that the ITM display increases to a level near that of the applied

calibration gas value.

b) Remove test gas and observe that the ITM display decreases to “0”.

Initial operational tests are complete. PI-700 VOC gas sensors are factory calibrated prior to shipment, and
should not require significant adjustment on start up. However, it is recommended that a complete calibration
test and adjustment be performed 16 to 24 hours after power-up.

Refer to zero and span calibration

instructions in Section 3.4.