Set autospan level, Table 2 maximum and minimum autospan settings – Detcon PI-700 User Manual
Page 24
Model PI-700
PI-700 Instruction Manual
Rev. 2.5
Page 20 of 54
When the status list sequence is complete, the ITM will revert to the “View Sensor Status” text scroll. The
user can either: 1) review list again by executing another 3-4 second hold, 2) move to another menu item by
executing a momentary hold over PGM1 or PGM2, or 3) return to Normal Operation via automatic timeout of
about 15 seconds (the display will scroll “View Sensor Status” 4 times and then return to Normal Operation).
3.5.2
Set AutoSpan Level
Set AutoSpan Level is used to set the span gas concentration level that is being used to calibrate the sensor.
This level is adjustable from 1% to 99% of range depending on the range of the sensor (see Table 2) . The
current setting can be viewed in View Program Status.
Table 2 Maximum and Minimum AutoSpan Settings
Sensor
Full Scale
Range
Minimum
AutoSpan
Level
Maximum
AutoSpan
Level
Adjustment
Increment
Sensor
Full Scale
Range
Minimum
AutoSpan
Level
Maximum
AutoSpan
Level
Adjustment
Increment
1.00
0.01
0.99
0.01
95
1
90
Note 1
2.00
0.01
1.99
0.01
100
1
95
Note 1
3.00
0.01
2.99
0.01
150
1
100
Note 1
4.00
0.01
3.99
0.01
200
1
150
Note 1
5.00
0.01
4.99
0.01
250
1
200
Note 1
6.00
0.01
5.99
0.01
300
1
250
Note 1
7.00
0.01
6.99
0.01
350
1
300
Note 1
8.00
0.01
7.99
0.01
400
1
350
Note 1
9.00
0.01
8.99
0.01
450
1
400
Note 1
10.0
1.0
9.9
0.1
500
1
450
Note 1
11.0
1.0
10.9
0.1
550
1
500
Note 1
12.0
1.0
11.9
0.1
600
1
550
Note 1
13.0
1.0
12.9
0.1
650
1
600
Note 1
14.0
1.0
13.9
0.1
700
1
650
Note 1
15.0
1.0
14.9
0.1
750
1
700
Note 1
16.0
1.0
15.9
0.1
800
1
750
Note 1
17.0
1.0
16.9
0.1
850
1
800
Note 1
18.0
1.0
17.9
0.1
900
1
850
Note 1
19.0
1.0
18.9
0.1
950
1
900
Note 1
20.0
1.0
19.9
0.1
1000
1
950
Note 1
21.0
1.0
20.9
0.1
1500
1
1000
Note 1
22.0
1.0
21.9
0.1
2000
1
1500
Note 1
23.0
1.0
22.9
0.1
2500
1
2000
Note 1
24.0
1.0
23.9
0.1
3000
1
2500
Note 1
25.0
1.0
24.9
0.1
3500
1
3000
Note 1
30
1
25
Note 1
4000
1
3500
Note 1
35
1
30
Note 1
4500
1
4000
Note 1
40
1
35
Note 1
5000
1
4500
Note 1
45
1
40
Note 1
5500
1
5000
Note 1
50
1
45
Note 1
6000
1
5500
Note 1
55
1
50
Note 1
6500
1
6000
Note 1
60
1
55
Note 1
7000
1
6500
Note 1
65
1
60
Note 1
7500
1
7000
Note 1
70
1
65
Note 1
8000
1
7500
Note 1
75
1
70
Note 1
8500
1
8000
Note 1
80
1
75
Note 1
9000
1
8500
Note 1
85
1
80
Note 1
9500
1
9000
Note 1
90
1
85
Note 1
10000
1
9500
Note 1
Note 1:
When the AutoSpan value is between 1 and 25, the adjustment increment is 1
When the AutoSpan value is between 26 and 100, the adjustment increment is 5