beautypg.com

Specifications – Aquametrix P/R60 Differential pH/ORP Sensors User Manual User Manual

Page 5

background image

P/R60 & P/R65

Page 5

2. SPECIFICATIONS

MEASURING RANGES

pH:

0 to 14.00 pH

ORP: R60: +1999 mV

R65:

0-1000 mV / +500 mV field

selectable

FLOW RATE

10 ft./sec max. Flow should be as low as
possible in low conductivity water and in
solutions with high suspended solids

WETTED MATERIALS

P60/P65: CPVC, kynar/ceramic, glass,
titanium palladium alloy and EPDM

R60/R65: CPVC, kynar/ceramic, glass,
titanium palladium alloy, EPDM, and platinum

TRANSMISSION DISTANCE

5-wire Probes: 900 meters (3000 ft.)

2-wire Probes: Limited only by cable
resistance and voltage of power supply

SENSITIVITY

pH: Less than 0.005 pH

ORP: Less than 0.5mV

STABILITY

0.03 pH / 2mV per day, non-cumulative

PRESSURE LIMIT

100 psig @ 65°C

40 psig @ 95°C

TEMPERATURE LIMITS

-5 to 95°C (23 to 203°F)

POWER SUPPLY LIMIT

24VDC ± 4V

PROBE CABLE

15 ft. (4.6 m)

AUTOMATIC TEMPERATURE
COMPENSATION

-5 to 95°C (23 to 203°F)

P65C

OUTPUT SPAN: 1.14mA / pH

OUTPUT OFFSET: 12mA @ 7pH
+1.14mA

MAX LOAD: 450 Ohms

R65C

OUTPUT SPAN:

0-1000mV: 1.6mA / 100mV

+500mV: 1.6mA / 100mV

OUTPUT OFFSET:

0-1000mV: 12mA @ 500mV
+1mA

+500mV: 12mA @ 0mV +1mA

MAX LOAD: 450 Ohms

This manual is related to the following products: