SHARK TX/P pH, ORP, Conductivity, Flow Transmitter
Technical Data
pH
ORP
Conductivity
Flow
Display
Front Panel: 4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED
Bar Graph. Inside Panel: 2 x 16 alpha-numeric LCD display
Measuring Range
pH: 0.01 to 14.00
Temp: 0 to 100°C
or 32° to 212°F
ORP: -1999 to +1999mV
(Dependent on sensor)
Temp: 0 to 100°C
or 32° to +212°F
MΩ/
cm
3
0 to 19.9
0.01
µS/
cm
3
0 to 2.00
0.01
0 to 20.0
0.1
0 to 200
0.1
0 to
2000
1.0
mS/
cm
3
0 to 20
10
0 to 200
50
Flow: 0 to 9999 with
selectable flow rate units
Volume: 0 to 9999 with
Auto Range Flow rate
units: Gallons (GP),
Cubic Feet (CF), Liters
(LP), Cubic Meters (CM),
custom by entering factor
related to Gallons.
Time units: Seconds (S),
Minutes (M), Hours (H)
Temperature
Compensation
Automatic or Manual
0 to 100°C (32° to +212°F)
Not required
Automatic or Manual
User selectable tempera-
ture compensation slope
0.0 to 10.0% / °C.
0 to 100°C (32° to 212°F)
Not required
Temperautre Unit
°C or °F
Not required
Temperature Sensor
User selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTD
Calibration modes
Auto-Calibration,
Manual Calibration,
Temperature Display
Manual Calibration
Temperature Calibration
Dry Calibration
Sample Calibration
Temperature Calibration
K factor input
Ambient Conditions
Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Sensor to Trans-
mitter Distance
Differential Sensor: 3000 ft
Combination Sensor:10 ft
300 ft
2000 ft
AnalogOutput
4 to 20 mA
Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale),max. load 800Ω
Memory Back-up
All user settings are retained indefinitely in memory (EEPROM)
Mechanical
SHARKTX Enclosure:NEMA 4X, 1/4 DIN, polycarbonate enclosure with two 1/2” conduit holes
SHARKTXP Enclosure:NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARKTX Mounting: Universal Mounting kit for surface, pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Sensor Input
Probe: -600 to +600 mV
Temp. Sensor: 0 to 9999 Ω
Probe: -1999 to
+1999 mV
Temp. Sensor:
0 to 9999 Ω
Cell: 0 to 9999 Ω
Temp. Sensor: 0 to 9999 Ω
Paddle: 0 to 2000 Hz
Invalid Entries
Invalid entries cannot be stored
Manual Test Mode
Process value can be simulated with arrow keys to verify correct setup of outputs
Output Hold
4 to 20 mA output is placed on hold when the transmitter is in Menu mode
Calibration Data
Recall data from last calibration, calibra-
tion mode, 1st&2nd accepted buffer value
and probe mV output, calibration tempera-
ture, calibration slope, and probe efficiency
Recall data from last calibra-
tion, calibration buffer ac-
cepted value and cell resistance,
calibration temperature.
Recall store K factor.
Auto Return
User selectable auto return if the transmitter is left in menu mode for more than 10 min.
Display Damping
User can select rate at which SHARK updates display. Enables display damping of unstable process
Net Weight
SHARK-TX: 0.71 lbs (0.32 kg)
SHARK-TXP: 0.25 lbs (0.12 kg)