Rice Lake Condec UMC600IS User Manual
Page 15
12
UMC 600IS Installation Manual
Parameter
Description
Values
1
Graduations
NTEP to 10,000 (80,000 available).
2
Display resolution
1, 2, 5, 10, 20, 50, 100
3
Decimal point
0, 0.0, 0.00, 0.000, 0.0000
4
Span select
Parameter 4 is not configurable: Span selection is automatically set for the
following ranges: Lo = 1.7 – 3.3 mV/V; Hi = 0.5 – 2.0 mV/V
5
Digital averaging
1, 2, 4, 8, 16, 32, A1 = 8-4-2; A2 = 16-8-2
6
Tare mode
ATNR, AUTO, FIXED, BOTH (inhibit with motion)
7
AZM band
Off, 0.5, 1, 3, 5, and 10 divisions
Use 0.5 for H-44, bench, counter and livestock applications; use 3 for vehicle,
axle-load, and railroad scales
8
AZM/PAZ aperture
±1.9%, 100% of capacity includes push-to-zero, H-44: 1.9%
9
Motion
Off, 1, 3, 5, 10, 20, 50 divisions H-44: vehicle, axle, livestock, RR, 3.0; all other 1.0
10
Displayed units
Lb and kg conversion
11
Setpoint mode
Not supported
12
Zero band
Not supported
13
Weigh mode
Not supported
14
Serial output
Port 1, demand, continuous, baud rate, G/T/N, or display (Port 2 not supported)
Table 3-1. Parameter Overview
Figure 3-4. Configuration Menu
XXXXXXX
XXXXXXX
XXXXXXX
GRADS
1
5
2
10
0.0
0.00
OPTION 5
OPTION 4
OPTION 3
OPTION 1
OPTION 2
SERIAL
CALIBR
CONFIG
RES
resolution
DECPT
decimal point
50
20
100
OPTION 6
100
selection
list,
see table 3-2
on page 15
0.000
0.0000
1
4
2
8
AVERAGE
digital average
A1=8-4-2
16
32
ATNR
AUTO
A2=16-8-2
FIXED
BOTH
OFF
0.5
1
3
5
10
1.9
FS
TARE
enable tare
AZM
band
AZM/PAZ
band
1
Off
3
10
5
20
kg
lb
lb. conversion
MOTION
band
DISPLAY
50
SETPOINT
MODE
Not
supported
ZERO BAND
MODE
OPTION 7
WEIGHING
MODE
SERIAL
serial output
port 1
demand
continous
baud rate
G/T/N
display
Not
supported
Not
supported