0 configuration, 1 digital configuration – Rice Lake Condec UMC600IS User Manual
Page 14
Configuration
11
3.0
Configuration
Prior to calibration, the UMC 600IS must be digitally
configured, or assigned a set of operating parameters.
The first three parameters are directly related to
calibration and must be set before proceeding to
calibration mode.
3.1
Digital Configuration
3.1.1
Parameter Overview
Table 3-1 on page 12 lists configuration parameters
and describes their values. The following paragraphs
give the procedure for configuring the UMC 600IS.
3.1.2
Configuration Procedure
1. Unscrew the two screws on the face plate
bracket (See Figure 3-1). The bracket drops
down, exposing four program switches on the
left.
Figure 3-1. Accessing the Program Switches
2. Temporarily remove the unit’s flexible black
display panel by gently pushing down and
lifting the panel up and out at its center to
expose the configuration and calibration
instructions printed on the surface below. The
switch function table defines the appropriate
front panel switch settings for the CONF and
CAL modes.
3. Close switch SW1-2, marked CONF (2), by
moving to the up position (see Figure 3-2). A
prompt appears with a parameter number and
data value.
The parameter identifier is a number, 1–14, that
correlates to the CONFIG chart on the upper left of
the switch map panel. Selected data represents the
value being entered into the unit configuration data.
For example, 1 100 sets the indicator to 10,000
graduations (see Table 3-2 on page 13).
When configuration is complete, set SW1-2 open
(down) to return the unit to normal operating mode.
Figure 3-2. Closing Switch 2
The function keys shown in the table in Figure 3-3
have alternate functions during digital configuration.
Figure 3-3. Front Panel Key Functions
NOTE:
The TARE RECALL key functions as a previous
screen key in CONFIG mode.
3.1.3
Digital Configuration Parameters
Table 3-2 on page 13 lists the configuration display
prompts (Prompt 1) and their value selections for
displayed graduations. Prior to calibration, the UMC
600IS must be digitally configured, or given its set of
operating parameters. The first three parameter
selections are directly related to calibration and must
be set up before proceeding to the calibration mode.
These parameters include the number of graduations,
the resolution, and the decimal point location in the
weight data, all of which define the scale capacity.
Table 3-3 lists Prompts 2 and 3 and Table 3-4 has
Prompts 5, 6, and 7. Prompts 8, 9, and 10 are shown in
Table 3-5 and Prompts 11 through 14 are listed in
Table 3-6 both on page 15.
1
6
2
7
3
8
4
9
5
0
CE
ENT
TARE
RECALL
ZERO
NET
GROSS
TARE
lb/kg
CONV
1 SET
POINT
2 SET
POINT
CAPACITY
600 x 0.1
ZERO
NET
GROSS
lb
kg
MOTION
3 4
2
1
OPEN
Zero
Net/Gross
Tare
lb/kg Conv
Parameter Select Parameter Data
Select
Subparameter
Select
Subparameter
Data Select
1
6
2
7
3
8
4
9
5
0
CE
ENT
TARE
RECALL
ZERO
NET
GROSS
TARE
lb/kg
CONV
1 SET
POINT
2 SET
POINT
CAPACITY
600 x 0.1
ZERO
NET
GROSS
lb
kg
MOTION
DEAD LOAD (3)
CONF (2)
CAL (1)
3 4
2
1
OPEN
NORM
Parameter Identifier
Selected Data
ZERO
NET
GROSS
TARE
lb/kg
CONV