beautypg.com

Menu tab, Alarm, Signature (/as1 option) – Yokogawa Button Operated DX2000 User Manual

Page 87: Burnout, rjc

background image

8

Installation and W

iring

Recommended Replacement

Periods for W

orn Parts

Setup Items and

Default V

alues

IM 04L42B01-02E

Menu tab

*2

Alarm

Setup Item

Selectable Range or Selections

Default Value

Setting

Basic setting

Reflash
Rate of change > Decrease
Rate of change > Increase
Indicator

On/Off/On-1s

*3

/On-2s

*3

1 to 32
1 to 32
Hold/Nonhold

Off
1
1
Nonhold

Switch, Relay

Internal Switch > AND
Relay > AND
Relay > Action
Relay > Hold

Relay > Relay Action on ACK

*1

None/S01/S01-S02/.../S01-S29/S01-S30
None/I01/I01-I02/.../I01-I35/I01-I36
Energize/De_energ
Hold/Nonehold
Normal/Reset

None
None
Energize
Nonhold
Normal

Hysteresis

Meas CH > High/Low
Meas CH > Delta High/Low
Math CH > High/Low
Ext. CH > High/Low

0.0 to 5.0
0.0 to 5.0
0.0 to 5.0
0.0 to 5.0

0.5
0.0
0.0
0.0

A/D, Memory

Setup Item

Selectable Range or Selections

Default Value

Setting

Without /AS1

Scan interval > Scan mode

Normal > Scan interval

Normal > A/D integrate
Fast

> Scan interval

Fast

> A/D integrate

Normal/Fast
125ms/250ms(DX2004, DX2008)

1s/2s/5s(DX2010, DX2020, DX2030, DX2040,

DX2048)
Auto/50Hz/60Hz/100ms
25ms (DX2004, DX2008)

125ms (DX2010, DX2020, DX2030, DX2040,

DX2048)
600Hz

Normal
125ms

1s

Auto
25ms

125ms

600Hz

Memory > Data kind

Display/E+D/Event

Display

With /AS1

*3

The Scan interval items are the same as on a DX without the /AS1 option.
Memory > Data kind

Display/Event

Display

†: You cannot select this option on a DX with external input channels (/MC1) or when the multi batch function (/BT2) is in use.

Signature (/AS1 option)

*3

Setup Item

Selectable Range or Selections

Default Value

Setting

Process type

Batch/Continue

Batch

Sign from recorder

Off/Sign1/Sign1+2/Sign1+2+3

Off

Signature at batch stop

On/Off

Off

FTP transfer at signing

On/Off

Off

Burnout, RJC

Setup Item

Selectable Range or Selections

Default Value

Setting

First-CH, Last-CH

Channel number

001

Burnout set > Mode

Off/Up/Down

Off

RJC > Mode
RJC > Volt

Internal/External
–20000 to 20000µV

Internal
0

Setup Items and Default Values

This manual is related to the following products: