9. Technical Information

9.7 Calibration Calculation

The PH72 meter is calibrated at 2 points using 2 standard solutions. The first calibration
is done so a certain line is drawn through the calibration point (Figure 9.9). The second
calibration is done so a line is drawn through the first and second calibration points
(Figure 9.10). One-point calibration is a simplified method where only the first
calibration of 2-point calibration is performed.

Note:

A calibration point is the emf corresponding to the pH value of a standard
solution used. (Refer to Table 4.1.)

pH0

pH14

F0909.EPS

pH7, 0mV

+mV

-mV

Calibration line

Theoretical line

Calibration point

Figure 9.9

Calibration at First Point

pH0

pH14

pH7, 0mV

+mV

-mV

Second
calibration point

Theoretical line

First
calibration point

Calibration line (slope)

Asymmetry potential

F0910.EPS

Figure 9.10

Calibration at Second Point