6-1

6. Maintenance

6. Maintenance

6.1 For Optimum Meter Performance

The Model PH72 Personal pH/ORP Meter is simple to operate, but is a precision
instrument. To ensure accurate results from the meter, the following precautions should
be observed.

Flow Diagram

Measurement

Maintenance

Calibration

Storage

F060101.EPS

Table 6.1

Precautions in Each Step

Calibration

Measurement

Maintenance

Storage

T0601.EPS

Calibration using standard solutions:

• 1- or 2-point calibration. 2-point calibration is recommended for

accurate pH measurement.

• Always use certified standard solutions (pH 2, 4, 7, 9, 10, or 12).

Sample solution requirements:

• pH range: 0 to 14 pH
• Temperature: 0 to 80

؇

C (0 to 100

؇

C when using needle type or test

tube size pH sensor)

• After measurement, rinse off remaining sample solution from the

electrode thoroughly.

• Avoid a place with high temperature and humidity.
• Keep the wetting cap (moisten the cotton wad in the cap with a few

drops of water) attached to prevent the glass electrode and liquid
junction from drying out.