3 sensor replacement, 4 rehydrating the glass electrode – Yokogawa PH72 Personal pH/ORP Meter User Manual
Page 52

IM 12B03D02-01E
6-3
6. Maintenance
F0602.EPS
Soak in 0.1 mol/l hydrochloric acid
0.1 mol/l dilute
hydrochloric acid
Cleaning water
After 10 to 20
minutes
Rinse thoroughly in water
Figure 6.2
Acid Cleaning
6.3 Sensor Replacement
Since a pH sensor undergoes chemical changes with time, its performance deteriorates
gradually. Under normal operating conditions a sensor can be used for one or two years.
Sensor deterioration, however, speeds up depending on the nature of a sample solution,
e.g., a high temperature solution. Storage conditions also affect the sensor life.
If
or
appears while calibration is being performed with a sensor for
which
is displayed, replace the sensor.
6.4 Rehydrating the Glass Electrode
A dry glass electrode gives fluctuating pH readings. If a glass electrode has dried up,
soak it in water (tap water) for 1 to 2 hours or longer to rehydrate. The sensor will give
stable pH readings.
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)