6. Maintenance

6.6 Cleaning and Drying Connectors

Deteriorated insulation between connector pins can cause inaccurate readings. To remove
stains and/or moisture that may cause deteriorated insulation, clean the connector with a
dry cloth or a cloth moistened with anhydrous alcohol. If necessary, use a dryer.

Wipe off stains and/or moisture on
meter connector with a dry cloth.

Check that there are no stains on
the O-ring.

O-ring

Hot air

Use a dryer if necessary to
remove moisture from
connector of sensor cable.

F060601.eps

CAUTION

Use anhydrous alcohol to clean the connectors so no moisture remains. Dry connectors
completely.