<8. Troubleshooting>

8-1

IM 11B06E01-01E

8. Troubleshooting

8.1 Install

Phenomena

Cause

Countermeasure

Fail to copy the data into the

system disk when installing.

Invalid of attribution of the

property of JEROS_AP

folder.

Un-click “Enable to archive

the folder” at the archive

attribution of JEROS_AP

folder.

8.2

Diagnostic window (Web browser)

Phenomina

Cause

Countermeasure

Diagnostic window can not

be opened.

No HTML file.

The copy of the file is not

created.

Re-install *1

Diagnostic window can

be opened but the check

window of the mapping file

does not come up.

Java application is not in

operation.

The file copying is not suc-

cess.

Re-install *1

No setting of Java is in

operation.

No license of Java is

installed.

Check the license of Java.

"TASK" is not OK on the

Modbus status window.

Modbus TCP Client is not in

operation.

No license of Modbus is

installed.

Check the license of Mod-

bus.

*1:

Re-install has to be executed after ASGW_MAP.CSV file is deleted.