Data -4 – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual
Page 33
<5. Diagnostic Window>
5-4
IM 11B06E01-01E
5.3
Checking the analyzer communication
status and displaying the mapping data
The communication status window is displayed when [Modbus I/F] - [comm.status] is clicked.
UPDATE:
The data is updated.
ID:
StnID (define in the mapping file)
TYPE:
GC8000/GC1000/PCAS/ASIU (defined in the mapping file)
IP Addr:
IP address (define in the mapping file)
TASK:
The status is “OK” when the communication task is in operation.
COMM:
The status is “NG” when the communication error occurs.
WRITE:
The status is “NG” when it fails the writing, but is “OK” when it successes.
READ OK:
Number of success in reading
READ NG:
Number of failure in reading
WRITE OK:
Number of success in writing
WRITE NG:
Number of failure in writing
The mapping data window is displayed when [Modbus I/F] - [mapped data] is clicked.
4th Edition : Nov. 22, 2011-00