<4. Engineering>
4-16
IM 11B06E01-01E
(3) Allocating DO information if the above values are not the same
If the values in (1) and (2) are not the same, the values of Device Label in DI/DO must be set to
those shown in DevLabelVar. Copy all values of Device Label in DevLabelVar and paste them to
the corresponding values of Device Label in DI/DO.
Copy:
Select the values of Device Label shown in (1). Then, right-click on the selected area
and choose Copy from the context menu.
Paste:
Select Channel 17 shown in (2), which is the first DO value of Device Label. Then,
right-click and choose Paste from the context menu.
Download: Send the replaced values of Device Label to ASGW by [File] – [Download].