Yokogawa EJX115A User Manual
Page 80
<5. Parameter Summary>
5-2
IM 01C25T01-06EN
Function
Label
Item
Contents
Default value Handling
*
1
Device
information
Chg universal rev
Change the HART universal
revision
HART 5 or HART 7
M
Country
HART 7
Country code
US, JP, DE, FR, ES, RU, CN JP
W
Dev id
Device ID
R
Distributor
Yokogawa
R
Drain vent matl
Drain and vent plug material
W
Extra No.
Customizaion number
R
Ext SW
External zeroing permission
Disabled or Enabled
Enabled
W
Fill fluid
Fill fluid
W
Final asmbly num
Final assembly number
W
Fld dev rev
Field device revision
R
Gasket matl
Gasket material
W
Isoltr matl
Capsule material
W
Mftr Date
Manufactured date
R
Model 1
Memo field for MS code 1
32 alphanumerics
W
Model 2
Memo field for MS code 2
32 alphanumerics
W
Model 3
Memo field for MS code 3
32 alphanumerics
W
Num of RS
Number of remote seal
W
Process Conn matl Process connection material
W
Process Conn size Process connection size
W
Process Conn type Process connection type
W
RS fill fluid
Fill fluid of remote seal
W
RS Isoltr matl
Remote seal material
W
RS type
Remote seal type
W
Serial No.
Serial number
R
Sofftware rev
Software revision
R
Style No.
Style number
Style number of product
R
Universal rev
Universal revision
16 alphanumerics
R
Cfg chng count
HART 7
Configuration change counter
0
R
Reset Cfg chng flag
HART 7
Reset Configuration change flag
M
Device Profile
HART 7
Device Profile
Process
automation
device
R
Max dev vars
HART 7
Max device variables
3
R
Model
Model
Model name + Measurement
span in the Suffix Codes
Ex) “EJX110 M”
R
Device
Variable
Simulation
Simulate
HART 7
Execution of device variable
simulation
Execute the simulation
M
Diag
Applicable
Diag Applicable
Appicable blockage detection
Disabled or Enabled
RG
Diag DPComp Diag DPComp
fDP compensation selection
Compensation or
Non-compensation
Compensation
WG
Diag Error
Diag Error
Results detected by ILBD or Heat
trace monitoring
RG
*1:
Handling:
R=Read only, W=Read & Write, M=Method, A=Applicable for option code AL, G=Applicable for option code DG6,
D=Applicable for differential pressure transmitters. Do not change these parameters for pressure transmitters.
*2:
Max three Burst Messages (Burst Message 1, Burst Message 2, Burst Message 3)
*3:
The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.