Vanguard DMOM‐600 User Manual
Page 30

Auto‐Ohm 200 S3 / DMOM‐200 S3 / DMOM‐600 USER’S MANUAL
REV 2
26
Press the
[START]
key to start the test.
h. The following screen will be displayed temporarily:
The screen will be updated as shown below while the current is being ramped up:
Once the current is ramped up to the selected test current, the initial test results will be
displayed and the burn‐in sequence will be initiated. The following screen will be
displayed:
The current will then be ramped down and then the final test results will be displayed as
shown below:
NOTES
A Pass/Fail ([P] or [F]) will be displayed if you had entered a passing
resistance range in step b.
If you had selected the automatic test mode in step c, you can now
disconnect the cables and re‐connect them as needed. A test will be
initiated automatically 2 seconds after the re‐connection is sensed.
TEST RESULTS
497
p
50.01 A
BURNING IN: 04
497
50.01 A
test in progress...
ramping current
12.30A 27
test in progress...
AUTO TEST MODE