beautypg.com

Vanguard DMOM‐600 User Manual

Page 13

background image

REV 2 Auto‐Ohm 200 S3 / DMOM‐200 S3 / DMOM‐600 USER’S MANUAL

9

 

 

1.5

Auto-Ohm 200 S3 Technical Specifications

Table 1. Auto-Ohm 200 S3 Technical Specifications

TYPE

Portable Micro-Ohmmeter

PHYSICAL SPECIFICATIONS

17" W x 7" H x 13" D (43.2cm x 17.8cm x 33cm); Weight: 16.8 lbs (7.6 Kg)

INPUT POWER 100-240 Vac, 14A, 50/60Hz

RESISTANCE READING

RANGE

1 micro-ohm to 10 milliohms at 200 A (0.1 micro-ohm resolution)
1 micro-ohm to 5 ohms at 1A

ACCURACY

1A to 4.99A: 1% ±10 micro-ohms
5A to 9.99A: 1% ±2 micro-ohms
10A to 200A: 1% ±1 micro-ohm

TEST CURRENT RANGE

1 Ampere to 200 Amperes (selectable in 1A steps);
Thermally protected DC power supply

DISPLAY

Backlit LCD (128 x 64 pixels), viewable in direct sunlight and low light levels

KEYPAD

Rugged, 44key "QWERTY" membrane keypad

INTERNAL TEST RECORD

STORAGE

128 test records. Each record can contain up to 64 readings

EXTERNAL TEST RECORD

STORAGE

Up to 999 test records on external USB Flash drive.

COMPUTER INTERFACE

RS-232C

PC SOFTWARE

Windows XP/Vista/7 Analysis Software (included with purchase)

SAFETY Designed to meet IEC 61010 (1995), UL 61010a, and CASC22.2 standards

ENVIRONMENT Operating: 10˚C to 50˚C (15˚F to +122˚F)

Storage: 30˚C to 70˚C (22˚F to +158˚F)

HUMIDITY (MAX) 90% RH @ 40˚ C (104˚ F) non-condensing

ALTITUDE (MAX) 2000m (6562 ft) to full safety specifications

CABLES 30 ft (9.1 m), #1 AWG test cables, power cord, ground cable

OPTIONS Shipping case, 15 ft test cables, C-Clamp set, Hand Spike Set, Dual Ground

option

WARRANTY One year on parts and labor

 

NOTE 

The above specifications are valid at nominal operating voltage and at a 
temperature of 25C (77F). Specifications may change without prior notice. 

This manual is related to the following products: