J. attenuator matching and final check – Teledyne LeCroy AP033 User Manual

Page 53

background image

Operator’s Manual

922260-00 Rev A

47

the boards into the housing being careful to align the screw holes in the end cover with the
corresponding channels in the corners of the housing.

7. Insert and tighten the two screws which secure the end panel to the ProBus interface housing.

Avoid over tightening the screws as the cover may warp.

8. Replace the four push button caps, pressing each fully to seat the cap on the button shaft.

J. Attenuator Matching and Final Check

1. Repeat the Performance Verification procedure to ensure compliance with the warranted

specifications.

2. Perform the Attenuator Matching Procedure listed on page 22. Apply calibration seals in

accordance with your quality procedures.

This concludes the Adjustment Procedure. Repeat the Performance Verification procedure to
complete the calibration of the AP033.