Performance Verification

This procedure can be used to verify the warranted characteristics of the AP033 Active Differential
Probe. You can do the performance verification without removing the instrument covers and
exposing yourself to hazardous voltages.

Adjustment should only be attempted if a parameter measured in the Performance Verification
Procedure is outside of the specified limits. Adjustment should only be performed by qualified
personnel.

The recommended calibration interval for the model AP033 Active Differential Probe is one year.
The complete performance verification procedure should be performed as the first step of annual
calibration. You can record test results on a photocopy of the Test Record at the end of this
section.

Test Equipment Required

Table 3 lists the test equipment and accessories (or their equivalents) that are required for
performance verification of the AP033 Active Differential Probe.

This procedure has been developed to minimize the number of calibrated test instruments
required. Only the parameters listed in boldface in the “Minimum Requirements” column must be
calibrated to the accuracy indicated.

Because the input and output connector types may vary on different brands and models of test
instruments, additional adapters or cables may be required.