beautypg.com

Foredom C.TXR-2 User Manual

Parts list, Foredom, C.txr−2 foot control, 230v

background image

Parts List

C.TXR−2
Foot Control, 230V

Form DN 12/10

FOREDOM

The Foredom Electric Company

16 Stony Hill Rd., Bethel, CT 06801

203−792−8622

Fax: 203−796−7861

www.foredom.net

}

Qty

Description

Part No.

Item

1

Cover

CP10852B

1

1

Trigger Switch, 250V

CP11027

2

2

Screw, #6−32 x 5/16, Slotted

CP10571

3

1

Fish Paper

CP10525

4

1

Circuit Board

CP10521

5

1

Base

CP10851

6

1

Front Base Pad

CP10859

7

1

Screw, #6−32 x 1"

CP10856

8

1

Washer, #6

CP10862

9

1

Rear Base Pad

CP10860

10

2

Strain Relief

CP10003−3

12

1

Male Cordset or Substitute Option

CP10541

11

1

Female Cordset

CP10868

13

1

Wire 3", Brown, Tinned

CP11163

14

1

Wire 3", Brown

CP11162

15

Tigger Switch

5

6

1

4
3

2

AC Mal

e P

lug

to

Power

Sourc

e

BLU

GRN/YLW

GRN

(G

RN
YL

W

)

BLK

(BRN)

BRN

BRN

Wiring Diagram

Insert adjustment screw (8) through washer (9) and into post

slot and threaded insert in cover (1). Tighten with screwdriver

until cover is just touching the trigger switch pushrod. Do not

tighten the screw further

to avoid pressing the Switch Pushrod

into a permanently "on" position.

Position wires around adjustment screw post so wires will not

interfere with cover travel.

Connect Wires to New Trigger Switch as shown in the

Wiring Diagram.

Push wires all the way in, so that no bare wire

is exposed. On Screw Terminal 1, push the wire lead all the way

into the socket then tighten the screw firmly. Push the other wire

lead all the way into terminal 6. Pull on all the wires, making sure

they are secure. Place Switch back into Base mounting bracket.

Snap cover (1) onto hinge posts on top of rear corners of base (6).

Disconnect Foot Control from Motor and Wall Outlet.

Instructions for Removing
and Replacing Trigger Switch

Unscrew cover adjustment screw (8).
Using flat blade of screwdriver inserted at Point A

between Cover (1) and Base (6), carefully pry Cover off.

5.

Remove Trigger Switch (2) from Base mounting bracket.

On older Switches, all the terminals will be push-in, spring clamp

type. On newer Switches, some terminals have screw fasteners.

To remove wires from the push-in terminals, insert 1.3mm [.05in]

approximate diameter pin or the blade of a jeweler's screwdriver

all the way into the terminal socket to release the spring clamp.

Grasp the tool and wire and pull them together out of the socket.

If you have difficulty removing the wires from the push-in terminals

and you have wire-cutting, stripping, and soldering tools, you can

cut the wires off where they exit from the terminals. To prepare

the cut wires for installation into the replacement Switch, strip

them back exactly 8.25 mm [5/16 in], then apply solder to the

stripped ends to fuse the strands together.

4.

3.

6.

7.
8.

2.

1.

Fe

mal
e Receptacle

to

Motor

For use with LX 230V Motors
and discontinued TX 230V Motors.

Cordset Option Part No. Cordset Descrption

CPOption20

CP10545 No Plug, Europe

CPOption21

CP10553 Male, United Kingdom

CPOption22

CP10554 Male, Continental Europe

CPOption25

CP10555 Male, India 10A

CPOption26

CP10556 Male, Australia

1

2

3

4

5

6

7

8

9

10

11

11

13

BRN

WHT

(BLU)

Point A

14

15