Adjusting the flow rate – SKC Limited EPAM-5000 User Manual
Page 65

5 - 7
Adjusting the Flow Rate
Introduction
The flow rate must be adjusted when it does not equal 4.0 LPM.
Adjusting the
flow rate
Follow the steps in the table below to adjust the flow rate.
1
Locate the adjustment screw on the front of the EPAM-5000.
2
Use the flow adjustment screw to adjust the flow rate.
To...
Turn the adjustment screw...
Decrease the flow rate
Counterclockwise
Increase the flow rate
Clockwise
3
Record the Flow Rate.
4
Detach the airflow calibrator and calibration airflow tubing.
5
Select enter to terminate battery status sensor and return to main
menu.
NOTE: Flow
meter must be
vertical when
reading
measurement.
Figure 5-2. Flow Adjustment Meter attached to EPAM-5000.