Sampling – SKC Limited EPAM-5000 User Manual
Page 34

Sampling
Introduction
Once you have selected a Particle Size and completed the Manual-Zero
process the EPAM-5000 is ready to begin sampling.
Conditions
The following conditions should be met before starting the sampling process.
Condition...
For further Information
See Page...
The correct particle size must be selected.
3-2
The correct sampling inlet must be attached.
3-2 –3-5
The date and time must be checked and/or set.
2-4
The Manual-Zero process must be complete.
3-7
The alarm level must be set if sampling with
the alarm feature.
2-5
Sampling
Follow the steps in the table below to begin the particle sampling.
Step Action
1
Turn unit on and press enter.
2
Select Run, and Choose Continue or Overwrite data.
3
Choose the memory storage type using the table below.
To... Select...
Erase all previously recorded
data points in all locations.
Overwrite, then
Select Yes to confirm,
Note: Selecting No will
cancel sampling process
without effecting memory.
Add data points to the next
consecutive location.
Continuation.
Note: See page 3-13 for explanation of location codes.
Continued on next page
3 - 9