Manual-zero – SKC Limited EPAM-5000 User Manual
Page 32

Manual-Zero
Introduction
Manual-Zero sets the measurement baseline of the EPAM-5000 to zero
mg/m
3
. The Manual-Zero check should take place prior to beginning a new
set of measurements.
Note: If using the Auto-Zero setting (default) the EPAM-5000 automatically
zeroes baseline every 30 minutes.
Manual-Zero
Follow the steps in the table below to Manual-Zero the EPAM-5000.
Note: The battery should be fully charged. Check battery status as described
in Chapter 5 Section 8.
Step Action
1
Be sure the appropriate sampling inlet is attached to the sensor
inlet of the EPAM-5000 using the table below.
If sampling…
Then insert the…
PM1.0 Particulates
1.0 impactor jet
PM2.5 Particulates
2.5 impactor
PM10 Particulates
10 impactor jet
TSP Particulates
Impactor sleeve only
Continued on next page
3 - 7