Performance Profi le
Time Display:
Time of day in hours and minutes (12 or 24-hour
clock) with AM and PM indicators
Volume Display:
Continually updated, based on corrected fl ow rate
multiplied by sampling time
Flow Fault:
If the pump is unable to compensate for longer than
15 seconds due to excessive back pressure, the pump
enters fl ow fault mode. During fl ow fault, the pump goes
into HOLD, the fault icon appears on the display during
the length of the fault, and the accumulated run time
display is frozen and retained. After 5 minutes in fl ow fault,
auto-restart is attempted every 5 minutes up to 10 times.
Power Supply:
Rechargeable 4.8-V nickel metal hydride (NiMH)
battery pack, 3.5-Ah capacity or
Rechargeable 4.8-V nickel cadmium (NiCad) battery
pack, 2.0-Ah capacity
A
battery eliminator is available (see Accessories);
use voids the UL Listing for intrinsic safety.
Charge Time:
6 to 8.5 hrs using PowerFlex charger
(varies with battery capacity
and level of discharge)
Size:
5.6 x 3 x 2.3 in (14.2 x 7.6 x 5.8 cm)
Weight:
22 oz (624 gm)
RFI/EMI Shielding:
RFI/EMI-shielded case, CE marked
Intrinsic Safety:
UL and cUL Listed
ATEX and ANZEx Models available
Tubing:
Requires 1/4-inch ID tubing
CE marked
UL Listed for intrinsic safety
Use only SKC-approved parts to ensure reliable performance. Failure