Airchek 2000 quick guide, Operation, Pump setup options – SKC 210-2000 AirChek 2000 User Manual

Page 3

background image

Security Code 
Must be pressed within 10 seconds
of previous command

Mode Change
Press [] to toggle between Run and Hold.

AirChek 2000 Quick Guide

Operation

Star Button
Scrolls through run time data, display options, and sampling parameters during pump setup

Up and Down Arrow Buttons 
Increase or decrease sampling parameters and toggle between display choices in setup

Button Sequence
 = press buttons individually
[] = press simultaneously when bracketed


= security code, always press in sequence

Pump Activation
Press any button.

Pump Setup Options

User Interface Level One—Set Up Flow Rate
To enter fl ow rate setup, pump must be in Hold. Press [
] to Run pump and press


within 10 seconds.

• Change and calibrate fl ow rate:
Flow rate and SET fl ash. Press  or  to change fl ow rate. Press  when fi nished; ADJ

displays and fl ashes. Press  or  to adjust the fl ow until pump and calibrator agree.
When fi nished, press  until End displays. Press [] to save new setting.

User Interface Level Two—Set Up Sampling and Display
To enter sampling and display setup, pump must be in Run. Press [
] to Hold pump
and press
 within 10 seconds. Select from the following options:

• Clear accumulated data:
Press

until CLr displays and then press []. Press  until End displays and then

press [].

• Change temperature scale (F/C):
Press

until temperature displays. Press  or  to switch units. Press  until End

displays and then press [] to save new setting.

• Change atmospheric pressure scale (mm Hg/ins Hg/millibars):
Press

until pressure displays. Press  or  to switch units. Press  until End

displays and then press [] to save new setting.

• Change time scale (12 Hr/24 Hr):
Press

until 12 Hr or 24 Hr displays. Press  or  to switch units. Press  until End

displays and then press [] to save new setting.

• Change time of day:
Press

until time displays. Press  or  to change fl ashing hour. Press  to move to

minutes and press  or  to change. Press  until End displays and then press []
to save new setting.

• Change sample time:
Press

until SГ 00 displays. Press  or  to change fl ashing digits. Press  until End

displays and then press [] to save new setting.

To exit without saving changes in User Interface Level Two, scroll to Esc and press [].