Airchek 2000 quick guide, Operation, Pump setup options – SKC 210-2000 AirChek 2000 User Manual
Page 3
Security Code
Must be pressed within 10 seconds
of previous command
Mode Change
Press [] to toggle between Run and Hold.
AirChek 2000 Quick Guide
Operation
Star Button
Scrolls through run time data, display options, and sampling parameters during pump setup
Up and Down Arrow Buttons
Increase or decrease sampling parameters and toggle between display choices in setup
Button Sequence
= press buttons individually
[] = press simultaneously when bracketed
= security code, always press in sequence
Pump Activation
Press any button.
Pump Setup Options
User Interface Level One—Set Up Flow Rate
To enter fl ow rate setup, pump must be in Hold. Press [] to Run pump and press
within 10 seconds.
• Change and calibrate fl ow rate:
Flow rate and SET fl ash. Press or to change fl ow rate. Press when fi nished; ADJ
displays and fl ashes. Press or to adjust the fl ow until pump and calibrator agree.
When fi nished, press until End displays. Press [] to save new setting.
User Interface Level Two—Set Up Sampling and Display
To enter sampling and display setup, pump must be in Run. Press [] to Hold pump
and press within 10 seconds. Select from the following options:
• Clear accumulated data:
Press
until CLr displays and then press []. Press until End displays and then
press [].
• Change temperature scale (F/C):
Press
until temperature displays. Press or to switch units. Press until End
displays and then press [] to save new setting.
• Change atmospheric pressure scale (mm Hg/ins Hg/millibars):
Press
until pressure displays. Press or to switch units. Press until End
displays and then press [] to save new setting.
• Change time scale (12 Hr/24 Hr):
Press
until 12 Hr or 24 Hr displays. Press or to switch units. Press until End
displays and then press [] to save new setting.
• Change time of day:
Press
until time displays. Press or to change fl ashing hour. Press to move to
minutes and press or to change. Press until End displays and then press []
to save new setting.
• Change sample time:
Press
until SГ 00 displays. Press or to change fl ashing digits. Press until End
displays and then press [] to save new setting.
To exit without saving changes in User Interface Level Two, scroll to Esc and press [].