Setup, Using user interface - level two – SKC 210-2000 AirChek 2000 User Manual

Page 15

background image

12

ADJ

Used during calibration with primary
standard calibrator (not for use with
CalChek feature
). Press  to increase
or  to decrease fl ow adjustment
until desired fl ow is indicated on an
att ached primary standard. Press  to
move to next parameter or the User Interface can be exited
by continuing to press  until End appears. Press [] to
save new fl ow and adjustment sett ings and exit the User
Interface. See page 17.

CAL

Use for CalChek calibration feature
only. Press [] to initiate single-
point calibration. Press  seven times
to start a full calibration. See CalChek
Calibration instructions on pages 19 to 23
.

Using User Interface - Level Two

Temperature

Press

 or  to toggle between

Fahrenheit (F) and Celsius (C). Press
 to move to next parameter.

Atmospheric
Pressure

Press

 or  to toggle through

inches of mercury (ins), millimeters
of mercury (mm), and millibars (m).
Press  to move to next parameter.

Sample Time
(SГ)

Program a specifi c run time. Press
 to increase or  to decrease the
time in minutes (from 1 to 999). Press
 to move to the next parameter.
Sample Time is not an option when a
DataTrac 2000 program is loaded into
pump memory (PROG). See page 14.

12-hr/24-hr
Clock and

Press  or  to toggle through

Delayed Start

standard (12-hour) time, military

(24-hour) time, and delayed start
(dELA). See Sett ing a Delayed Start
on page 15.
Press  to move to
next parameter.

Setup