beautypg.com

Calibration, Calchek error chart, Single-point calibration errors – SKC 210-2000 AirChek 2000 User Manual

Page 26: Full (multiple-point) calibration errors

background image

23

Calibration

CalChek Error Chart

Single-point Calibration Errors

Error Problem

Troubleshooting

E41

Correction required too large. A mismatch

of greater than 360 ml/min between the

flow setting on the pump and the reading

generated by the Defender Calibrator has

occurred.

Perform a full calibration. If this

fails, contact SKC Technical

Support at 724-941-9701 (800-

752-8472 [U.S. only]).

E48

Could not get a successful single point

calibration within fi ve fl ow readings.

Try the calibration again. If problem

persists, perform a full calibration.

Full (Multiple-point) Calibration Errors

Error Problem

Troubleshooting

E44

First fl ow reading greater than 750 ml/

min. The pump is fl owing faster than it

should, even though the calibration routine

delivered only a very small voltage to the

pump.

Check pressure sensor tubing to

ensure it is not pinched or blocked,

or contact SKC Technical Support

at 724-941-9701 (800-752-8472

[U.S. only]).

E45

Pump unable to achieve fl ow rate of 3000

ml/min possibly due to a blocked fl ow tube

or an air leak inside the pump.

Check pump’s fl ow tube to ensure

it is not blocked, or contact SKC

Technical Support at 724-941-9701

(800-752-8472 [U.S. only]).

E46

Analysis error in the data (rare).

Try full calibration again. If problem

persists, contact SKC Technical

Support at 724-941-9701 (800-

752-8472 [U.S. only]).

E47

Less than two bars appear in the battery

icon on the pump display indicating that the

battery is too low. There must be at least

two bars to begin a full calibration.

Recharge the battery.

No Code

At conclusion of full calibration, pump does

not verify to within 5%.

Pump not at ambient conditions for

at least two hours. Retry calibration

after pump has been at ambient

conditions for two hours.

Pump not running for fi ve minutes

prior to calibration. Run pump for

fi ve minutes and retry calibration.

Errors That Can Occur During Both Calibration Modes

Error Problem

Troubleshooting

E42

Unstable average. There is too much

variation in the fl ow readings.

Try the calibration again. If problem

persists, contact SKC Technical

Support at 724-941-9701 (800-

752-8472 [U.S. only]).

E43

Serial time out. The calibrator is not

communicating with the pump.

Check cable connection. If loose or

disconnected, connect properly.