Performance profi le – SKC 210-2000 AirChek 2000 User Manual
Page 5
2
Performance Profi le
Flow Range:
1000 to 3250 ml/min
(5 to 500 ml/min requires optional low fl ow adapter kit)
Compensation Range:
3000 ml/min at 15 inches water back pressure
2000 ml/min at 30 inches water back pressure
1000 ml/min at 30 inches water back pressure
Accuracies: Timing:
1 min/month at 25 C
Atmospheric
Pressure:
± 0.3 ins Hg
Flow Rate:
± 5% of set-point after calibration
Battery Charge
Level Indicator:
Icon displays at full, mid, and low charge (see page 6)
Temperature Range:
Operating:
32 to 113 F (0 to 45 C)
Charging:
50 to 113 F (10 to 45 C)
Storage:
-4 to 113 F (-20 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Protect sample pump from weather when in use outdoors.
Altitude:
Do not use pump beyond 7500 ft
Run Time:
NiCad Battery: 10 hrs minimum at above compensa-
tion ranges; dependent on media used. See Table 1.
NiMH Battery: 12 hrs minimum at 2000 ml/min and up
to 30 inches water back pressure; dependent on media
used. See Table 2.
Battery Eliminator: Pump provides extended runs.
Sample Time (SГ)
AirChek 2000 sample time can be set up to 999 minutes
Setting:
from the pump keypad or up to 43,200 minutes (30
days) from a PC using DataTrac 2000 Software. During
sampling, the pump LCD will display remaining run time
(will count time down from a set number of minutes)
when sample time is set from the keypad. The pump
LCD will show elapsed time in minutes when the pump
is programmed from a PC. If run time is longer than
9999 minutes (6.8 days), the sample time displayed on
the LCD will roll over to 1 after reaching 9999.
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min)
1.0
1.5
2.0
2.5
3.0
Filter/Pore Size (μm)
25-mm MCE, 0.8
6
9
12
15
18
25-mm MCE, 0.45
14
22
28
35
40
37-mm MCE, 0.8
2
3
4
5
6
37-mm PVC, 5.0
1
1
2
2
2.5