Setting a Delayed Start
The Delayed Start feature is available in pump version 2.59 or higher.
When sett ing the pump manually for sampling to begin within a 12-hour
timeframe, follow this procedure:
1. With the pump running, press [] to place the pump in HOLD and
enter within 10 seconds. You are in User Interface Level Two.
2. Press
until
SГ
appears. Press or to set a sample time. This is
required to set a delayed start.
3. Press
until the display reaches the 12-hr/24-hr clock. Press or
until the display shows a fl ashing dELA (delayed start).
4. Press
. Time will display with hours fl ashing. Press to increase or
to decrease the hour. Press to move to minutes. Press to increase
or to decrease the minutes.
The time entered here will be the next occurrence of this time. There
is no a.m. or p.m. designation.
5. Press
until End appears.
6. Press
[].
7. The PROG icon and a fl ashing HOLD will appear in the upper left
corner of the display. The pump is now set for delayed start.
If sett ing a delayed start using DataTrac 2000 Soft ware, refer to the
DataTrac 2000 Operating Instructions (included on the soft ware CD).
A programmed delayed start can be viewed in the DataTrac 2000
Scheduler.
Deleting a Delayed Start
This is the same method used to delete a DataTrac 2000 program from the
pump (see Deleting a DataTrac 2000 Program on page 16).
1. With the pump running (will only run in short bursts), press [] to
place the pump in HOLD and enter within 10 seconds. You are
in User Interface Level Two.
2. Press
to scroll to the fl ashing OFF indicator and press [].
3. Press
until End displays.
4. Press
[]. The PROG icon should be gone. If not, repeat process until
it disappears.
Setup