23

ITEM

FUNCTION

Loop test

This item invokes a method that performs a test on the primary 4-20ma current loop

Damping

Shows, and allows the user to set, the damping constant for the primary 4-20ma
current loop

D/A trim

This item invokes method that performs the D/A trimming of the primary 4-20ma
current loop

Aux Loop Cfg Menu

This menu is used to access the secondary 4-20ma loop calibration procedures. It contains the following items:

ITEM

FUNCTION

SV is

Shows, and allows the user to set, the variable assigned to the secondary 4-20ma
current loop

Loop test

This item invokes a method that performs a test on the secondary 4-20ma current loop

Damping

Shows, and allows the user to set, the damping constant for the secondary 4-20ma
current loop

D/A trim

This item invokes a method that performs the D/A trimming of the secondary 4-20ma
current loop

SV is is one of the following:

SV is

MEANING

Level

Level

Head Temp

Head temperature (if available)

Not Assigned

Blank line