Calibrate menu, Loop config menu, Low reference menu – Ronan X96S LEVEL GAGE User Manual
Page 26: High calibrate menu
22
ITEM
FUNCTION
Ref Mode
Shows, and allows the user to set, the reference/calibrate mode
Ref Time
Shows, and allows the user to set, the number of seconds of data to collect for a
reference or calibrate sample
MinRefCnts
Shows, and allows the user to set, the minimum raw value to use for a reference or
calibrate sample
Ref Mode is one of the following:
Ref Mode
MEANING
Empty/Full
Vessel will be Empty (air) for reference and Full (filled with process) for calibration in
measuring area.
Process
Process material in measuring area (not necessarily empty and full for reference and
calibration). User will supply actual levels during reference and calibration.
Absorber
Absorber placed in radiation path.
Calibrate Menu
This menu is used to access the various calibration procedures. It contains the following items:
ITEM
FUNCTION
Low Reference
Selecting this item takes the user to the Low Reference menu
High Calibrate
Selecting this item takes the user to the High Calibrate menu
Clear Ref/Cal
This item invokes method that clears the level reference
Low Reference Menu
This menu is used to perform the low reference procedure. It contains the following items:
ITEM
FUNCTION
Reference
This item invokes a method that performs the low reference procedure
Ref Level
Shows, and allows the user to set, the reference level value
Ref Cap
Shows the raw captured reference counts
High Calibrate Menu
This menu is used to perform the high calibrate procedure. It contains the following items:
ITEM
FUNCTION
Calibrate
This item invokes a method that performs the high calibrate procedure
Cal Level
Shows, and allows the user to set, the calibrate level value
Cal Cap
Shows the raw captured calibrate counts
Loop Config Menu
This menu is used to access the primary 4-20ma loop calibration procedures. It contains the following items: