Alarms
The
Alarms
menu is used to configure the parameters associated with the analog alarms.
ITEM
FUNCTION
Source
Show, and allows the user to set the source
Alarm Type
Shows, and allows the user to set the alarm type
Setpoint
Shows, and allows the user to set the alarm set point
Setpoint2
Shows, and allows the user to set the second alarm set point
4
Hysterisis
Shows, and allows the user to set the alarm hystersis percent
Source is one of the following
Alarm Type
MEANING
Level
Uses the Level for the source of the alarm
Head Temp
Uses the Head Temperature of the detector for the source of the alarm
Filtered Counts
Uses the Filtered Counts from the detector for the source of the alarm
Alarm Type is one of the following:
Alarm Type
MEANING
None
Alarm not yet set
Low
Alarm when the source is equal to or lower than Setpoint
High
Alarm when the source is equal to or higher than Setpoint
Range
Alarm when the source is equal to or lower than Setpoint OR the source is equal
to or higher than Setpoint2
Hardware Menu
The Hardware menu is used to define the type of hardware used to provide measurements and radiation. It contains the
following items:
ITEM
FUNCTION
System Hardware
Shows the user to a list of the hardware modules in the system and the status of these
modules
Source Type
Selecting this item takes the user to the Source Type menu
Analog Out Cnfg
Shows and allows the user to set the where the source of power is internal or external
System Hardware Menu
The System Hardware menu shows the user to a list of the hardware modules in the system and the status of these modules:
4
The second alarm set point is only used when the alarm type is range.