Filtering menu – Ronan X96S LEVEL GAGE User Manual

Page 17

background image

13

ITEM

FUNCTION

Filtering

Selecting this choice takes the user to the Filtering menu

Empty Clamp

Selecting this choice takes the user to the Empty Clamp menu

Detector Fault

Selecting this choice takes the user to the Detector Fault menu

Linearization

Selecting this choice takes the user to the Linearization menu

Scan Time

Shows the amount of time to accumulate each level sample and allows the user to
change the time value.

Filtering Menu

The Filtering menu is used to configure the parameters associated with the level measurement filter. It contains the following
items:

ITEM

FUNCTION

Dyn Track

Shows the current state of the dynamic tracking filter (enabled or disabled) and allows
the user to change the state.

Threshold

Shows the maximum number of raw counts (for scintillation) or raw analog value (for
ion chamber) that the input can vary from the expected value without considering that
count/value potentially erroneous. Also allows user to change this number.

Walk Avg

Shows the number elements to maintain and average in the walking average filter.
Also allows user to change this number.

Slope Smpls

Shows the number measurements to make between each recalculation of the density
slope. Also allows user to change this number.

Noise Filter

Shows the maximum number of potentially erroneous measurements in a row to
bridge before deciding that a step change has occurred in the level value. Also allows
user to change this number.

Monitor

Shows the current state of the filtering mechanism.

Monitor (filter state) one of the following:

Monitor

MEANING

ERROR

Filter is not initialized (this state should not occur during normal operation of the
X96S Level Gage)

FILL

The walking average buffer is filling.

TRACK

The walking average buffer is filled and the filter is tracking changes in the level value

REFILL

A step change has occurred and the walking average buffer is refilling.