Digital inputs menu – Ronan X96S LEVEL GAGE User Manual
Page 24
20
ITEM
FUNCTION
Setpoint
Shows, and allows the user to set, the alarm set point
Setpoint2
Shows, and allows the user to set, the second alarm set point
6
Hysterisis
Shows, and allows the user to set, the alarm hysterisis percent
Polarity
Shows, and allows the user to set, the alarm polarity
Alarm Source is one of the following:
Alarm Source
MEANING
Level
Operate this TTL output when Level is in alarm as defined by the Alarm Type and set
points.
HeadTemp
Operate this TTL output when the head temperature is in alarm as defined by the
Alarm Type and set points
System Alarm
Operate this TTL output when the X96S detectes a problem
Detector Flt
Operate this TTL outpu twhen a there is a problem with detector
Rad Disc
Operate this TTL output when the Rad. Disc. function triggers.
Auto Cal Ref
Operate this TTL output when the X96S is performing an automatic calibration
Auto Cal Err
Operate this TTL output if the X96S detects an error while performing an automatic
calibration
Not Used
This TTL output is not currently in use.
Alarm Type is one of the following:
Alarm Type
MEANING
None
Alarm not yet set
Low
Alarm when the source is equal to or lower than Setpoint
High
Alarm when the source is equal to or higher than Setpoint
Range
Alarm when the source is equal to or lower than Setpoint or equal to or higher
than Setpoint2
Polarity is one of the following:
Polarity
MEANING
Not Driven
Normally not driven
Driven
Normally driven
Digital Inputs Menu
This menu is used to view and configure the digital inputs. It contains the following item:
ITEM
FUNCTION
Input 1
Selecting this item takes the user to the Input 1 menu
Input 2
Selecting this item takes the user to the Input 2 menu
Input 3
Selecting this item takes the user to the Input 3 menu
Input 4
Selecting this item takes the user to the Input 4 menu
Input 5
Selecting this item takes the user to the Input 5 menu
Input 6
Selecting this item takes the user to the Input 6 menu
Input 7
Selecting this item takes the user to the Input 7 menu
Input 8
Selecting this item takes the user to the Input 8 menu
6
The second alarm set point is only used when the alarm type is range.