15

ITEM

FUNCTION

Table Entry #

Shows, and allows the user to select, an entry in the linearization table

Entry Used

Shows if the entry is used or not.

Measured

Shows, and allows the user to set, the measured value associated with this linearization
table entry. This is a value calculated by the X96S.

Actual

Shows, and allows the user to set, the actual value associated with this linearization
table entry. This value is the result of actual level knowledge, and compares to the
Measured value above.

Set Entry

This item invokes a method that sets a table entry

Remove Entry

This item invokes a method that removes a table entry

Level Config Menu

The Level Config menu is used to configure the parameters associated with the level measurement. It contains the following
items:

ITEM

FUNCTION

Units

Shows, and allows the user to set, the level units used

Low Range

Shows, and allows the user to set, the level value to be mapped to 4ma on the current
loop output, if level is selected to control that current loop.

High Range

Shows, and allows the user to set, the level value to be mapped to 20ma on the current
loop output, if level is selected to control that current loop.

Units is one of the following:

Units

MEANING

ft

feet

m

meter

in

inch

cm

centimeter

mm

millimeter

Head Temp Config Menu

The Head Temp Config menu is used to configure the parameters associated with the detector electronics temperature
measurement. This function is used primarily in high-temperature applications where the temperature exceeds the electronics
temperature specifications. It contains the following items:

ITEM

FUNCTION

Temp Units

Shows, and allows the user to set, the units to be used for head temperature

Low Range

Shows, and allows the user to set, the temperature value to be mapped to 4ma on the
current loop output, if head temperature is selected to control that current loop.

High Range

Shows, and allows the user to set, the temperature value to be mapped to20ma on the
current loop output, if head temperature is selected to control that current loop.