I/o menu descriptions, continued – Fairbanks 2800 Series Intrinsically Safe Instrument User Manual
Page 41
Section 3: Programming
01/12
41
50647 Rev. 9
3.5.1. I/O MENU DESCRIPTIONS, CONTINUED
PROMPT
COMMENT(S)
LoSEt
Sets the weight value associated with 4mA.
Usually
"
0
".
1. At
the
LoSEt
prompt, press
ENTER
.
2. Use the numeric keypad to enter desired weight value.
3. Press
ENTER
.
hiSEt
Associated weight value of 20MA.
Usually the capacity of the scale.
1. Press
ENTER
at the
hiSEt
prompt.
2. Use the numeric keypad to enter the desired weight value.
3. Press
ENTER
.
4 AdJ
Press
ENTER
to display a current number representing the 4mA Output.
─ See note on key functions.
20 AdJ
Press
ENTER
to display a current number representing the 20mA
Output.
─ See note on key functions.
AgtESt
LP4
- TO -
LP22
.
Press
ENTER
to perform an analog test.
4mA thru 22mA will be output momentarily in 2mA steps.
When adjusting either 4mA or 20mA, the following numeric keys increase
or decrease the reference value in the fashion that follows.
Increased by:
Decreased by:
9
10,000
4
10,000
8
1,000
3
1,000
7
100
2
100
6
10
1
10
5
1
0
1
ConoPt
Used for Continuous Output.
Choices include the following:
3052r
FbStd
FbPUtr
troniC
ConCon
ProtoA
Protocol setting for Port A choices include selections for the following:
bAUd 9600, 600
,
1200
,
2400
, and
4800
ChAr 8
and
7
PAritY nonE
,
odd
, and
EuEn