Accessories, continued – Fairbanks 2800 Series Intrinsically Safe Instrument User Manual
Page 13
Section 1: General Information
01/12
13
50647 Rev. 9
1.3. ACCESSORIES, CONTINUED
1.3.8. ACCESSORY 2860 SAFE AREA DATA CONVERTER
(
20523
)
This accessory is service programmable and configurable for RS232 or 20mA
data outputs.
This accessory can also be configured to provide a 4-20 ma analog output.
Enclosure: NEMA 4; Fiberglass
4 -20 ma Specifications: 16 bit resolution (+/- .01 integral linearity)
Current loop voltage compliance: 7vdc minimum to 40vdc maximum (typical
voltage = 24vdc)
Full scale settling time: 8msecs
Output impedance: 25 meg
Alarm current: 3.5 to 24 mA (underload/overload conditions)
─ Offset @ 25 degrees C; +/- .1% of full scale
Offset drift: +/- 25ppm of full scale per degree C
Total output error: 20mA @ 25 degrees C: +/- .2% of full scale max
Total output drift: +/- 50ppm of full scale per degree C-max
1.3.9. ACCESSORY 2875 INTRINSICALLY SAFE
INTERCONNECTION CABLE (
21737
)
This is the only approved cable for interconnecting the Accessory 2802, and
for connecting to the Model 2800 to the Accessory 2850.
Specifications: 2 pair - 16AWG; 0.48" O.D.
C A U T I O N
The 4-20ma accessory is a passive device!
The power is to be supplied by others, and MUST
BE
a
DEDICATED
and an
ISOLATED
power supply.