Using an External Computer to Control the Test Set
Using National Instruments VISA
If you are using National Instruments VISA, be sure to set the following
variables as follows:
VI_ATTR_SEND_END = VI_FALSE ‘
This specifies whether to assert END
during the transfer of the last byte of the buffer
VI_ATTR_TERMCHAR = 0x0A ‘
This is the termination character. When
the termination character is read and
VI_ATTR_TERMCHAR_EN
is enabled
during a read operation, the read operation terminates.
VI_ATTR_TERMCHAR_EN = VI_TRUE ‘
This is a flag that determines
whether the read operation should terminate when a termination
character is received.
VI_ATTR_SUPPRESS_END_EN = VI_FALSE ‘
Specifies whether to
suppress the
END
bit termination. If this attribute is set to
VI_TRUE
, the
END
bit does not terminate read operations. If this attribute is set to
VI_FALSE
, the
END
bit terminates read operations.
Write Commands:
Append all commands with “\n,” for example,
*rst\n
.
Read Commands:
The Z5623A Option H48 returns data terminated by
\r\n
.