Experiment 1: Introduction to the Operation of the
Dissolved O

2

Sensor

Purpose

The purpose of the experiment is to explore the basic operation of the Dissolved Oxygen
Sensor: how to set it up, how to calibrate it, and how to obtain the most accurate
measurements with it.

Materials and Equipment Needed

Dissolved Oxygen Sensor (PS-2108)
600 ml beaker
PASCO ScienceWorkshop
400 ml deionized water saturated with air
PASPort Interface and DataStudio software
Reference tables for dissolved oxygen
aquarium pump (optional)

Optional Equipment Suggested

To saturate deionized water with air, fill a clean container one-third full with deionized water,
seal it, and shake vigorously for 10 seconds. Alternatively, bubble air through the deionized
water for 15 minutes using an aquarium pump.

Procedure

1. Set up and calibrate the Dissolved Oxygen Sensor in DataStudio

according to the calibration procedure detailed in this manual.

2. Place the probe in the test solution and stir it gently for thirty

seconds.

3. Click the Start button on the DataStudio toolbar to begin recording

data (Data Run #1).

4. Being careful to keep the stainless steel band submerged to at least

1 cm under the surface of the water, gently stir the probe in the test
solution until the dissolved O

2

readings stabilize.

5. Click the Stop button on the DataStudio toolbar to end data

recording.

Figure 1
Gently stir the probe in the
test solution