PASPort Dissolved Oxygen Sensor

Model No. PS-2108

8

®

2. Use the supplied plunger to push the membrane cartridge

out of the cartridge housing (Figure 7A).

3. Examine the O-ring (Figure 8) and replace it if it is

damaged (See "Replacing the O-ring" below.).

4. Insert a replacement membrane cartridge and use the

plunger to push it down until it is seated at the end of the
housing that has the O-ring (Figure 7B).

5. Fill with electrolyte and reassemble the probe as directed

in steps 4 through 7 of "Changing the Electrolyte
Solution."

Replacing the O-ring

The O-ring should rarely if ever require replacing. However, if it develops nicks or splits and
begins to allow leakage of the electrolyte solution from the probe, replace it as follows:

1. Follow steps 1and 2 of the procedure for replacing the membrane

cartridge (above).

2. After removing the membrane cartridge, remove the O-ring with a pair

of fine-tipped tweezers and insert a new O-ring.

3. Insert the membrane cartridge as directed in step 4 of “Replacing the

Membrane”.

4. Fill with electrolyte and reassemble the probe as directed in steps 4

through 7 of "Changing the Electrolyte Solution."

Figure 7
Removing and replacing the
membrane cartridge.

A

B

Figure 8
Examine the O-ring for
damage.