8.2.10
Calibrate CH1 0V
Submenu Group
8.2.11
Calibrate CH1 10V
Submenu Group
Calibrate ch1 0V
Vin=TB1-2 GND=TB1-4
Calibrate ch1 0V
0 CALIBRATING ——
Calibrate ch1 0V
*** DONE ***
Calibrate ch1 0V
Iin=TB1-2 GND=TB1-4
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
Calibrate ch1 10V
0 CALIBRATING ——
Calibrate ch1 10V
*** DONE ***
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
Advance to
Calibrate ch2 0V
To Calibrate: Connect Voltage Source (+) TB1-2, (-) TB1-4.
Input 0V and press Enter.
This message is displayed during calibration.
This message is displayed when the 0V calibration is fin-
ished.
The display will automatically return to the Calibrate CH1 0V
submenu. Press the Down arrow key to advance to the CH1
10V calibration.
To Calibrate: Connect Voltage Source (+) TB1-2, (-) TB1-4.
Input 10V and press Enter.
This message is displayed during calibration.
This message is displayed when the 10V calibration is fin-
ished.
The display will automatically return to the Calibrate CH1 10V
submenu. Press the Down arrow key to advance to the CH2
0V calibration.