beautypg.com

Luminex 100E Hardware User Manual User Manual

Page 18

background image

Luminex 100E User Manual

x

MAP Technology

3 - 6

PN 89-00002-00-013 Rev. D

The power input connector type is a Amphenol JTP02RE-22-
21P(014), and it connects with Amphenol JT06RE-22-21S(014).
Equivalent parts can be used. The Luminex 100E analyzer operates
off of five separate dc voltage levels, including 24 volts, 15 volts, -15
volts, and 6 volts. The pinout follows below:

+15

pins E, F

-15

pin G

+24

pin J

+6

pin C

+15 Return

pins P, S

+24 Return

pin H

+6 Return

pin D

Chassis ground

pin A

5.4 volts

pins L, M, W

5.4 Return

pins F, K, N

5.4 V (+) sense

pin B

5.4 V (-) sense

pin R

-15 volts return

pin U

The user must provide power interruption and fusing means.

The power required from the host system to run the Luminex 100E
analyzer is listed below. The voltages listed are required at the dc
power connector. This power requirement allows for the use of
additional components at a later date.

Note:

The 6 volt line must be at least 5.5 volts.

Note:

The +15 volt supply must be at a minimum of 90% of rated

output at least 15 (±5) milliseconds before the 6 volt supply reaches
10% of rated output to ensure proper operation.

Note:

The laser supply voltage (+5.4 volts) must be at a minimum of

90% of rated output no later than 15 (±5) milliseconds after the 15
volt supply is at 90% of rated output. The 5.4 volt supply must be
capable of providing a voltage of 5.3 to 5.5 volts dc at the sense
points located within the Luminex 100E analyzer. The load initially
varies between 0 and 50 W. After 15 seconds, the load decreases to
less than 40W.