Physical dimensions, excluding mounting brackets, connectors,
and tubing, are: 31.1 cm (12.25 inches) W x 50 cm (19.7 inches)
D x 29.2 cm (11.5 inches) H
•
Weight: 21.8 kg (48.0 lbs.)
•
Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to +50°C and up to
80% noncondensing
•
Underwriters Laboratories, Inc. (UL
®
) installation category: UL
Installation Category II, as defined in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree II, as defined in Section
3.7.3.2 of UL 61010A-1
•
Classification of microspheres > 80%
•
Misclassification of microspheres < 0.5%
•
Internal sample carry over < 3%
•
Soluble background fluorescence emission at 575 nm
automatically subtracted from median fluorescence intensity
(MFI) value
Sensitivity
•
Detect 1000 fluorochromes phycoerythrin (PE) per microsphere
•
Reporter channel dynamic range: 3.5 decades of detection
Capacity
•
Sample core: 15 - 20 µm core at 1 µL/sec sample injection rate
Optics
•
Reporter laser: 532 nm, nominal output 15-16.5 mW, maximum
of 500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
•
Classification laser: 635 nm, 9.1 mW ± 6%, maximum output
25 mW, diode; mode of operation, continuous wave (CW)
•
Reporter detector: Photomultiplier tube, detection bandwidth of
565 - 585 nm
•
Classification detector: Avalanche photo diode (APD) with
temperature compensation
•
Doublet discrimination detector: Avalanche photo diode (APD)
with temperature compensation