Maintenance

Lamp replacement

Scale & Units

Scale: In the automatic Scaling mode, the y-axis is
automatically adjusted so that the total scan is displayed.
The manual Scaling mode allows sections of the scan to be
displayed.
Units: Choice of absorbance or transmittance.

Send Data

To send Data to a printer, computer or USB memory stick
(Type A)

Reading Mode

Single Reading: One measurement result is shown after
tapping on Read.
Carousel 1 inch square: Optional measurement of carousel
insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.

Recall
measurement
data

Recall saved measurement data, wave-length scans or time
courses, see

Save, recall, send and delete data, page 154

.

Instrument Setup Basic data of the instrument.

Table 6 Time course options (Continued)

Options

Description

C A U T I O N

Potential Chemical, Biological Eye and Skin Hazards.
Only qualified personnel should conduct the tasks described in this section of the
manual.

N O T I C E

Remove all cells remaining in the instrument. Dispose of the cells or their
contents in an authorised manner.

D A N G E R

Risk of electric shock
Disconnect the instrument from the power source before the lamp replacement
procedure is commenced and until the lamp replacement procedure is complete.

C A U T I O N

Burn hazard.
Wait until the lamp cools down. Contact with the hot lamp can cause burns.

N O T I C E

Only hold lamp at lamp holder. Avoid touching the glass, as substances on the
skin can bake onto the lamp bulb and thus accelerate the ageing process of the
lamp.