Hach-Lange DR 6000 Basic User Manual User Manual
Page 162

English 162
Wavelength Scan mode – recording of absorbance
and transmission spectrums
In the Wavelength Scan mode, the absorbance of the light in a solution
over a defined wavelength spectrum is measured.
The reading results can be displayed as a curve, as percentage
transmittance (%T) or as Absorbance (Abs). The collected data can be
printed as a table or a curve.
The data is available for formatting changes. These include automatic
scaling and zoom functions. Maximum and minimum values are
determined and shown as a table.
The cursor can be moved to any point on the curve for the purpose of
reading off the absorbance or transmittance value and the wavelength.
The data associated with each data point can also be shown as a table.
Reading Mode
Single Reading: A single reading is displayed after Read is
pressed.
Carousel 1 inch square: Optional measurement of carousel
insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.
Recall
measurement
data
Recall saved measurement data, wavelength scans or time
courses, see
Save, recall, send and delete data, page 154
.
Instrument Setup
mode
Basic data of the instrument, refer to
(absorbance, concentration and transmittance readings),
page 159
.
Table 4 Multi wavelength setup options (Continued)
Options
Description