Hach-Lange DR 6000 Basic User Manual User Manual
Page 160

English 160
Table 3 Options for Single Wavelength
Options
Description
More
For further Options
Recall
measurement data
symbol
Recall saved measurement data, wavelength scans or time
courses, refer to
Save, recall, send and delete data,
page 154
.
% Trans/Abs
Switches from % transmittance to concentration or
absorbance in reading mode.
λ
Wavelength
To input the wavelength setting. Use the alphanumeric
keypad to enter the reading wavelengths. A wavelength in
the range of 190–1100 nm can be entered.
Timer icon
This functions as a stopwatch. It helps to make sure that
the steps of an analysis are correctly timed (e. g. reaction
times, wait times, etc., can be exactly specified). When the
specified time has elapsed, a sound is emitted. The use of
the timer has no influence on the reading program.
Concentration
Factor: Off/On
Multiplication factor for converting absorbance values into
concentration values.
Concentration
Resolution
To select the number of decimal places.
Save as User
Program
To store the selected parameters as a User Program, refer
to
Single Wavelength (absorbance, concentration and
transmittance readings), page 159
Reading Mode
Single Reading: One measurement result is shown after
tapping on Read.
Continuous Readings: After the zero measurement, all
readings are displayed automatically and continuously.
Carousel 1 inch square: Optional measurement of
carousel insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.
Recall
measurement data
Recall saved measurement data, wavelength scans or time
courses, refer to
Save, recall, send and delete data,
page 154
.
Instrument Setup
mode
Basic data of the instrument, refer to
Save, recall, send and
delete data, page 154
.
Table 3 Options for Single Wavelength (Continued)
Options
Description