beautypg.com

Command line interface commands – Fluid Components International FS10A User Manual

Page 30

background image

FS10 Series

28

Fluid Components International LLC

Command Line Interface Commands

INFO

Read version information (Hardware, Firmware, Date code, other values (#1.. #45). [The values in items

1 through 45 are listed]

MEAS

Read most recent flow measurement values (#220..#238). [The values in items 220 through 238 are listed]

RCFG B

Read configuration parameters (#80..#132) from bank B (0..9). [Example: *RCFG 7 (to read bank 7 parameters)]

SAVE B

Save currently active parameters to bank B (0..9) (Level 2 – Field password required). [Example: *SAVE 7]

RCL B

Recall parameters from bank B as the active parameters. (Level 2 – Field password required). [Example: *RCL 7]

PASSWD NNNN

To enter the password for either Level 1 (Factory) or Level 2 (Field). [Example *PASSWD 19113]

EXIT

To exit the current password level (undo having entered a password). The current parameters will automatically be

saved to Bank 0 upon exit.

NNN=

Update item #NNN (1..238) in the “active” (RAM) parameters with value. Value may be either integer or floating

point (scientific notation). [Example 1: *228=1.100119E+03 enters value 1.100119E+03 into item 228, Example 2:

*1=002 enters 002 into item1, Example 3: *110=800 enters 800 into item 110]

B:NNN

Read item #NNN (1..238) in Bank B (0..9). For items #1..#45, and items #220..#238 the Bank is ignored. If no bank is

specified, the currently active (RAM) parameters will be accessed. [Example 1: *7:85 returns item 85 from bank 7 in

this form: 7:85>5.053665E-02, Example 2: *8:119 returns item 119 from bank 8 in this form: 8:119=030]