Burkert Type 8072 User Manual
Page 13
11
Technicaldata
Measurement range
• 8032
• 8072
• 0.3 m/s to 10 m/s,
0.3 m/s corrresponds to a flow
rate of 0.5 l/min in a DN06
fitting
• 2 to 1200 l/min, with a vis-
cosity > 5 cps;
3 to 616 l/min, with a viscosity
< 5 cps;
Accuracy
• 8032 (see Fig. 7)
• 8072
• ±1 % of the full scale
1)
, with
on site calibration or when
using the teach-in feature
or
± (1 % of the full scale
1)
+ 3%
of the measured value), when
using the standard K factor
• ±1% of the measured value
Repeatability
• 8032
• 8072
• ±0.4% of the measured value
• ±0.03% of the measured
value
Linearity (8032 only)
±0.5 % of the full scale
1)
Measuring element
paddle-wheel of the S030 fitting
or oval gears of the S070 fitting
1)
Full scale = 10 m/s
Type of fluid
depends on the fitting used
(refer to the operating instruc-
tions of the fitting used)
Fluid viscosity
depends on the fitting used
(refer to the operating instruc-
tions of the fitting used)
Rate of solid particles (8032
only)
max. 1%
Fluid temperature
• with fitting S030 in PVC
• with fitting S030 in PP
• with fitting S030 in stainless
steel, PVDF, or brass
• with S070 fitting, housing in
aluminium
• with S070 fitting, housing in
stainless steel
• 0 to 50 °C (see Fig. 8)
• 0 to 80 °C (see Fig. 8)
• -15 to 100 °C (see Fig. 8)
• max. 80 °C
• max. 100 °C
Fluid pressure
• 8032 with S030 fitting in
metal
• 8032 with S030 fitting in
plastic
• 8072 with fitting S070 with
flanges
• 8072 with any fitting S070
except S070 with flanges
• PN16 max. (see Fig. 8)
• PN10 max. (see Fig. 8)
• PN10 max.
• PN55 max.
English
Technical data