Brooks, Model 5851 i – Brooks Instrument 5851i User Manual

Page 13

background image

2-3

Installation and Operation Manual
X-TMF-5851i-MFC-eng
Part Number: 541B109AAG
September, 2009

Brooks

®

Model 5851

i

Section 2 Installation

Recommended installation procedures:

a. The Model 5851

i should be located in a clean dry atmosphere rela-

tively free from shock and vibration.

b. Leave sufficient room for access to the electrical components.
c. Install in such a manner that permits easy removal if the instrument

requires cleaning.

d. The Model 5851

i Mass Flow Controller can be installed in any

position. However, mounting orientations other than the original
factory calibration (see data sheet) will result in a ±0.5% maximum
full scale shift after re-zeroing.

e. When installing controllers, be aware that sharp abrupt angles in the

system piping directly upstream of the controller may cause a small
shift in accuracy. If possible, have at least ten pipe diameters of
straight tubing upstream of the 5851

i mass flow controller.

NOTE: The control valve in the Model 5851

i provides precision control

and is not designed for positive shut-off. If positive shut-off is required, it is
recommended that a separate shut-off valve be installed in-line.

Figure 2-1 Model 5851

i Dimensions

Connection

"X" Dim.

1/4" Compression

5.70

Fitting

145

1/4" Tube VCO

5.22

133

1/4" Tube VCR

5.56

141

3/8" Compression

5.82

Fitting

148

1/2" Tube VCO

5.48

139

1/2" Tube VCR

5.86

149