Brooks, Model 5851 i – Brooks Instrument 5851i User Manual
Page 13
2-3
Installation and Operation Manual
X-TMF-5851i-MFC-eng
Part Number: 541B109AAG
September, 2009
Brooks
®
Model 5851
i
Section 2 Installation
Recommended installation procedures:
a. The Model 5851
i should be located in a clean dry atmosphere rela-
tively free from shock and vibration.
b. Leave sufficient room for access to the electrical components.
c. Install in such a manner that permits easy removal if the instrument
requires cleaning.
d. The Model 5851
i Mass Flow Controller can be installed in any
position. However, mounting orientations other than the original
factory calibration (see data sheet) will result in a ±0.5% maximum
full scale shift after re-zeroing.
e. When installing controllers, be aware that sharp abrupt angles in the
system piping directly upstream of the controller may cause a small
shift in accuracy. If possible, have at least ten pipe diameters of
straight tubing upstream of the 5851
i mass flow controller.
NOTE: The control valve in the Model 5851
i provides precision control
and is not designed for positive shut-off. If positive shut-off is required, it is
recommended that a separate shut-off valve be installed in-line.