Section 4 Maintenance

4-6

Installation and Operation Manual

X-CM-QmB-eng

Part Number: 541B029AAG

June, 2011

QmB Series IP40, IP66, IP66XP

4. Provide an electrical signal to the Quantim representing a setpoint

of 50% of full scale flow and adjust the inlet and outlet pressures to
calibration conditions.

5. Verify that the electrical output signal reaches 100% of flow. For

MFC applications vary the setpoint over the 1% to 100% range and
verify that the output signal matches the setpoint.
a. If possible, connect a flow measurement device (such as a

rotometer) to the Quantim outlet to monitor the actual flow
behavior.

6. If the Quantim is functioning correctly, the problem may lie

elsewhere in the flow system. Re-verify the installation as well as the
upstream/downstream fluid system configuration.

7. Please contact Brooks Instrument for further assistance in

troubleshooting. Refer to the back cover of this manual for contact
information.

PRESSURE

GAUGE

PRESSURE

GAUGE

FLUID

SUPPLY

PRESSURE

PRESSURE

REGULATOR

QmB

FLOW

OPTIONAL

MEASURING

MEASURING

DEVICE

SHUT-OFF

VALVE

SHUT-OFF

SHUT-OFF

VALVE

OR

BACK

PRESSURE

REGULATOR

Brooks Instrument

TM

PRECISION MASS FLOW

Figure 4-1 Bench Troubleshooting Circuit